Troubleshooting guide / Error messages

56

9. Troubleshooting guide / Error messages

!

+

Whenever the microprocessor electronics registers a failure, a complete shutdown

of the circulator is performed. The alarm light "

" illuminates and a continuous

signal tone sounds.

• The circulator is operated without bath liquid, or the liquid level is insufficient.

Replenish the bath tank with the bath liquid.

• Tube breakage has occured (insufficient filling level due to excessive bath

liquid pumped out). Replace the tubing and replenish the bath tank with the
bath liquid.

• The float is defect (e. g., because damaged in transit). Repair by authorized

JULABO service personnel.

• Failure of one or both temperature sensors of the CU-unit.

• The cooling water temperature value lies below 5 °C.

• Cable of the working temperature sensor interrupted or short-circuited.

• Defect of the working or safety temperature sensor.

Working temperature and safety sensors report a temperature difference of
more than 100 K.

• Other errors (I

2

C-BUS errors)

• Error in A/D converter.

• Safety sensor defect.

The safety temperature value lies below the working temperature setpoint.
Set the safety temperature to a higher value.

• External control selected, but external Pt100 sensor not connected.

• Error in the measuring system of the external temperature sensor „EXT“.

• Error at the internal Interface

• Safety sensor 1:

Cable of the safety sensor 1 interrupted or short-circuited.

• Safety sensor 2:

Cable of the safety sensor 2 interrupted or short-circuited.

• Safety sensor 3:

Cable of the safety sensor 3 interrupted or short-circuited.

• Safety sensor in the cooling zone:

Cable of the safety sensor interrupted or short-circuited.