Troubleshooting

54

I

When the tank is full, open the front cover (this may require removal
of the final filter) to go into STANDBY mode.

Let the system stand for 1 hour for effective bacteria elimination.

J

(After 1 hour) Check that the Tank Outlet Tubing is secured into the

drain.
Open the Tank Outlet Valve to drain all the water from the tank (J).

K

Close the Tank Outlet Valve (K).

L

Close the front cover (L).

(this may require removal of the final filter)

The system will now go into FILLING TANK mode. Let the tank level fill
up to the 100% level.

Repeat steps I, J, K and L:

Open the front cover (this may require to remove the final filter)

to go into STANDBY mode (I).

Check that the Tank Outlet Tubing is secured into the drain. Open

the Tank Outlet Valve to drain all the water from the tank (J).

Close the Tank Outlet Valve (K).

Close the front cover (L). The system will now go into FILLING

TANK mode.

The tank is now sanitized.