Electrical requirements – Labconco PuriCare Vertical Flow Stations 3830020 User Manual

Page 11

Advertising
background image

7

Product Service 1-800-522-7658

Chapter 2: Prerequisites

Electrical Requirements


The different PuriCare models have the following
electrical requirements:


Model #

Requirements

38300- 00-09

115 VAC, 60 Hz, 12 Amps

38300- 20-29

230 VAC, 50 Hz, 6 Amps


All PuriCare Stations with model numbers ending in –
00 to –09 are designed for operation at 115 volts, 60Hz,
alternating current. PuriCare Stations with model
numbers ending in –20 to –29 are designed for
operation at 230 +/- 20 volts, 50 Hz alternating current.
A dedicated outlet with a circuit breaker rated at 20
Amps for 115 volt models and 10 Amps for 230 volt
models, should be located within 10 feet of the station.

Always follow the plug
manufacturer’s instructions for the
proper assembly and testing of the
plug and power cord.


THE ELECTRICAL OUTLETS OF
THE 115 VOLT PURICARE IS
PROTECTED BY A GROUND FAULT
INTERRUPTER CIRCUIT (GFIC).
LABCONCO DOES NOT
RECOMMEND INSTALLING A GFIC
ON THE OUTLET THAT THE
PURICARE PLUGS INTO.

!

Table 2-1