Turning on the System

Turning on the System

Note: Depending upon local electrical codes,
an uninterruptible power system (UPS) may be
required in your laboratory. A UPS is not
provided with the
Opsys MW™ system.

1. Connect

the

Opsys MW™ to the laboratory electrical supply

outlet.

CAUTION: The Opsys MW™ instrument must
be connected to properly grounded electrical
outlets. Obtain assistance from a qualified
electrician to verify that your electrical outlets
are properly grounded.

Before connecting the power cable, be sure
that the components have been connected to
each other as outlined in the previous section.

2. Push the instrument power switch once. After a series of self-

tests, the software version is displayed momentarily

Opsys MW. v1.0

before the main menu is displayed:

1: - - - - - - - -

SETUP PROGRAM

Note: The self-test results indicate whether
the system is operating properly. If a self-test
failure is reported, it is recommended that you
call Technical Service.

Dashes are initially displayed after the wash
protocol number (i.e., 1: - - - - - - - - in the
above example) if a wash protocol has not yet
been named.