Technical specifications


Communications

RS232 via a standard modem cable terminated with male and female 9-pin Sub-D connector at each
end. Cable length <3 metre. A USB- RS232 adaptor may be used.

4-20 mA output for flux

4-20 mA output for pipe temperature


Measurements
Logged to memory:

Detector

Range

Resolution

Accuracy

Flux (AT-F probe)

0 to 2400 pl/cm

2

/s

1 pl/cm

2

/s

10% at cal. value

Pipe temp

-40

C to +615

C

1

C

2

C at cal. value

Internal temp

-40

C to +125

C

1

C

1

C at cal. value

Flow

0 to 60 ml/min

0.1 ml/min

5% at cal. value


4-20mA output:

Note Range output range is variables from 1 to 2400pl/cm

2

/s and is selected using the Data

Download Package.


Memory
The memory will record approximately 18 months of data. Once memory is full data is logged in a first in first
out rolling cycle by memory sector.

Diagnostics

Operating temp

-40

C to 60

C

H2 sensor

Electrolytic conductivity

Flow

30 ml/min +/-3%

Power alarm

Insufficient supply voltage

Visual Alarm

Via red LED on instrument

Transmitted alarm

3.5 mA on 4-20 mA outputs

Log to memory


Operating lifetime

Pump

10,000 hr cumulative use


Probe Dimensions

Probe diameter

144 mm

Probe length

10 m (30 ft)

Probe head length

320 mm inc. allowance for 90

bend

Probe conduit diameter

16 mm

Probe conduit bend radius

35 mm


Probe fixing
Banding for >4” up to 32”
Magnets for large diameter pipes where banding not practical. (max working temperature 800

F 426

C).

Note; insulation and cladding are highly recommended and are essential if weather protection is required.
EMC. To achieve the immunity detailed above all wiring should be screened twisted pairs.

Detector

Range

Resolution

Accuracy

4 mA

20 mA

10mA

Flux (AT-F probe)

0 pl/cm

2

/s

2400pl/cm

2

/s *

1 pl/cm

2

/s

10% at cal. Value

Pipe temp

-40

C

500

C

2

C

2

C at cal value