Xylem photoLab 6100 VIS User Manual
Page 127

photoLab
®
6100 VIS
Operation
127
ba75847e01
08/2009
General AQA2
settings
Note
When the mode (Weeks or Measurements) is changed, all AQA2 intervals
are reset to the preset values.
Locking the method
Here you configure whether or not a method will be locked against
measurement if there is no valid AQA2 check or the interval for the AQA2
check has expired.
1
Select and confirm Mode.
The Mode selection field pops up.
2
Select and confirm Weeks or
Measurements.
AQA2 is active. For all methods,
the AQA2 intervals are entered
either in weeks or number of
measurements.
3
Accept the general settings with
[Apply].
<HOME>
[AQA]
– AQA2 setup
AQA2 setup
04/16/07 9:52
Mode
Weeks
Lock methods
Yes
Method...
Method list
1
Select and confirm Lock methods.
2
Select and confirm Yes.
The method lock is enabled.
Each method will be locked if the
AQA2 check is invalid or the AQA2
interval has expired.
<HOME>
[AQA]
– AQA2 setup
AQA2 setup
04/16/07 9:52
Mode
Weeks
Lock methods
Yes
Method...
Method list