:VOLT:OFFS 2.5V
:VOLT:OFFS MAX
Query
Syntax:
[:SOURce]:VOLTage:OFFSet?[<ws>MINimum|MAXimum]
Examples:
:VOLT:OFFS?
:VOLT:OFFS? MAX
Response:
NR2
Considerations:
1) The MAXimum offset is dependent on the amplitude.
2) The MAX and MIN arguments should not be used in a program message containing an AMPLitude command,
since these values are evaluated during parsing, based on the current value of the amplitude.
4.13.1.4 Clock Reference Source
:SOURce:REFerence:SOURce <clock source>
This command is used to select the source of the arbitrary waveform clock. This clock sets the arbitrary waveform point rate.
Arguments
Type:
Character
Options:
INTernal, EXTernal
Command Type: Setting or Query
Setting
Syntax:
[:SOURce]: REFerence
:SOURce<ws><option>
Examples:
:REF:SOUR INT
:REF:SOUR EXT
Query
Syntax:
[:SOURce]: REFerence :SOURce?
Response:
INT|EXT
4.13.1.5 Function
:SOURce:FUNCtion [:SHAPe]
The function command is used to set the type of waveform to be generated by the instrument.
Command Type: Setting or Query
Setting
Syntax:
[:SOURce]:FUNCtion[:SHAPe]<WS><OPTION>
Examples: : FUNC SIN
:FUNC ARB
Query
Syntax:
[:SOURce]:FUNCtion[:SHAPe]?
Examples:
:FUNC?
Response:
SIN|TRI|SQU|ARB|PUL
Considerations:
The following functions are available:
SINusoid, SQUare, TRIangle, ARBitrary, PULse