Section 5, Performance check procedures, 1 introduction – B&K Precision 4079 - Manual User Manual
Page 97: 2 test equipment

91
Section 5
Performance Check Procedures
5.1 Introduction
This section provides the procedure for checking the electrical performance requirements of the Model
4076 and 4079 Arbitrary Waveform Generators, as listed in instrument Operating Manual Section 1: “
Specifications”.
If the waveform generator fails to meet these checks, then you should perform the adjustment
procedure. Adjustment procedures are available on a separate document.
The Performance Check Procedures are identical for both channels on the Model 4079 and must be
performed separately for each channel.
At the end of this document, a table is provided to record the test results.
5.2 Test Equipment
The following table lists the equipment necessary to perform the performance tests. Any equivalent
equipment may be substituted for the recommended model.
Description
Minimum Requirements
Recommended Model
Oscilloscope
Bandwidth > 400MHz
Tektronix DPO4054
Distortion Analyzer
Frequency range 10Hz-100KHz Tektronix AA5001
Digital Voltmeter
6 1/2 digits, true RMS
Fluke 8845A
Universal Counter
200MHz, Time A B, Period
Pendulum CNT-90
Function Generator
10MHz output
BK Model 4045
50Ω Feed-through
±1% to 1GHz (high frequency)
Tektronix 011-0049-02
Precision 50Ω Feed-through
±0.1% (DC to 100KHz)
Tektronix 012-0482-XX
50Ω Coaxial Cables, with BNC connectors, 1 m,