B&K Precision 4079 - Manual User Manual
Page 99

93
Amplitude
setting
Minimum reading RMS Maximum reading RMS DVM reading RMS
10Vp-p
3.493V
3.578V
5Vp-p
1.743V
1.792V
3Vp-p
1.043V
1.078V
1Vp-p
0.343V
0.364V
100mVp-p
31mV
40mV
50mVp-p
14mV
22mV
5. Set the unit to generate a SQUARE wave with 1KHz frequency.
6
. CHECK that the measured voltages on the DVM at 10Vp-p, 5Vp-p, 3Vp-p, 1Vp-p, 100 mVp-p and
50mVp-p are in the accuracy range:
Amplitude
setting
Minimum reading RMS Maximum reading RMS DVM reading RMS
10Vp-p
4.940V
5.060V
5Vp-p
2.465V
2.535V
3Vp-p
1.475V
1.525V
1Vp-p
0.485V
0.515V
100mVp-p
44mV
56mV
50mVp-p
20mV
31mV
4 Set the unit to the ARBITRARY function and generate a SINE wave signal:
- Press the WAVE key
- Select F5 – ARB
- Select F4 – EDIT
- Select F3 – PREDEF
- Select F2 – FROM, press key 1 and ENTER
- Select F2 – DATA, press key 0 and ENTER
- Select F3 – LENG, press 1000 and ENTER
- Select F4 – SCALE, press 100 and ENTER
- Select F5 – EXEC
- Select F3 – YES
- Select PARAM
- Select F1 – FREQ, press 1000 Hz