1 – calibration event log, 2 – configuration event log – Liquid Controls Sponsler T675 - Cryogenic System Register User Manual

Page 29

Advertising
background image

Sponsler, Inc.

Model T675 Flow Totalizer

Technical Information

Pg. 26

3.3.2.1.1 – Calibration event log
Allows viewing the changes for the sealable calibration parameters in chronological order beginning with
the most recent. The last 30 Kfactor or linearizer changes are recorded. When full, the audit trail shifts the
oldest date and time stamp off Display. Changes made within the linearizer table or the linearizer status
are recorded as a single entry to the audit trail.

**** Calibration event log, Page 01 ****

006 01-Nov-07 16:33 –Factor = 148.914
005 02-Nov-07 19:34 –Factor = 149.345
004 04-May-06 14:25 –Lin. OFF
003 15-Mar-06 10:45 –Lin. table
002 23-Jan-06 17:42 –Lin. ON
001 15-Oct-05 14:52 –Factor = 150.115
Empty
Empty
Empty
Empty

Press any key to continue


Each page of the audit trail calibration event log contains 10 entries. The minimum log capacity is 30
events. The number of pages required to document this depends on the number of multiple changes that
occur per event. Any unused entry is displayed as Empty. Multiple changes executed during the same
Adjustment mode entry will be recorded as a single event and increment the calibration counter 1.

3.3.2.1.2 – Configuration event log
Allows viewing the changes for the sealable configuration parameters in chronological order beginning
with the most recent. The last 40 product, digits to right of decimal or unit of measure changes, or
temperature compensation enable/disable cycles are recorded. When full, the audit trail shifts the oldest
date and time stamp off Display.

*** Configuration event log, Page 01 ***

005 20-Sep-07 19:51 –DP = 2
004 09-Jun-07 20:34 –Tcomp ON
003 05-Feb-07 11:36 –Product =LOX
002 18-Aug-07 09:45 –New E.U. = POUNDS
001 31-Mar-07 10:18 –Tcomp Off
Empty
Empty
Empty
Empty
Empty

Press any key to continue


Each page of the audit trail configuration event log contains 10 entries. The minimum log capacity is 40
events. The number of pages required to document this depends on the number of multiple changes that
occur per event. Any unused entry is displayed as Empty. Multiple changes executed during the same
Adjustment mode entry will be recorded as a single event and increment the configuration counter 1.

Advertising