6 & 7, – spare, 8 – activate, 9 – loaner – Liquid Controls Sponsler T675 - Cryogenic System Register User Manual

Page 36: 4 – update software, 5 – hardware test/calibrate, 6 – restart system, 7 – serial echo/debug

Advertising
background image

Sponsler, Inc.

Model T675 Flow Totalizer

Technical Information

Pg. 33

3.3.2.6 & 7, – Spare
Available for future expansion.

3.3.2.8 – Activate
Permits stocking of the T675 by the customer without consuming the warranty period. Activating the T675
operating system automatically inserts the date of activation into the date of manufacture for warranty purposes.
If activation is required an activation code must be acquired from Sponsler Inc.

****** Activation required (0=no) ******

Current setting = 0



Enter setting (ESC to exit)?

3.3.2.9 – Loaner
A factory Privileged setting, establishes the allotted loan period the T675 will operate.

*** Loaner expiration DDMMYY (0=off) ***

Current setting = 0



Enter setting (ESC to exit)?

3.3.2.0 – Return to page 1
Returns to 675 Settings, Page1 menu.

3.4 – UPDATE SOFTWARE

Activates the system's InfraRed communications port. Once activated, data and system programs can be transferred
between the T675 and an IBM-PC. ** I.R. communication activated ** is displayed and an audible tone occurs every
second. Depressing the RESET key exits this function and reverts to the 675 Initialization options Display. This option is
disabled when the T675-PEM is installed.

3.5 – HARDWARE TEST/CALIBRATE

This function is PRIVILEGED passcode protected. Factory calibration of the A/D converter channels, cold starting a virgin
system, and running system diagnostics are performed in this function. A complete instruction sequence is contained in
the T675 Technical Addendum (MN-T675-TA).

3.6 – RESTART SYSTEM

Initiates system hardware reset. “

*SYSTEM RESET IN PROGRESS*

“ will be displayed prior to the automatic return

of the system to the Operating mode.

3.7 – SERIAL ECHO/DEBUG

Factory technician’s aid to confirm data transfer functions. “

** ECHO TEST ACTIVATED **

” will be displayed.

Depressing the Reset key exits this function and reverts to the 675 Initialization options Display.