Luminex® flexmap 3d® hardware, Recommended additional equipment, Luminex – Luminex FLEXMAP 3D User Manual User Manual

Page 31: Flexmap 3d, Hardware

Advertising
background image

CAUTION:

Luminex calibrators and verifiers contain sodium azide. Sodium
azide is toxic and may also react with lead and copper plumbing
to form highly explosive metal azides. On disposal, flush drains
with a generous amount of cold water to prevent azide build-up.
Consult the manual guideline Safety Management No. CDC-22,
Decontamination of Laboratory Sink Drains to remove Azide
Salts
(Centers for Disease Control, Atlanta, Georgia, April 30,
1976).

Luminex

®

FLEXMAP 3D

®

Hardware

The Luminex

®

FLEXMAP 3D

®

system includes the following hardware:

• The FLEXMAP 3D

®

instrument

• Personal computer (PC) and accessories

• Stereo speakers

• Power cables

• Three sample probes

• Off-plate reagent area

• 96 and 384 heater blocks

• Empty cubitainer for waste

• Sheath fluid intake line

• Waste fluid outtake line

• USB communication cable

• Barcode reader (optional)

• Sample probe height alignment kit and 384-well plate Probe Height Adjustment tool

• Monitor arm

NOTE: For information on calibrating the probe correctly, refer to the

FLEXMAP 3D 384-Well Plate Probe Height Adjustment Tool

section.

Recommended Additional Equipment

Successful operation of the Luminex

®

FLEXMAP 3D

®

system may require additional

equipment.

Uninterruptible Power Supply (UPS) or Surge Protector

Luminex recommends using either an uninterruptible power supply (UPS) or a surge
protector to protect your system from power outages. Use a UPS that provides 1300 watts for
at least 45 minutes. Select a surge protector that fits your requirements with regard to
electrical environment, endurance, suppressed voltage rating, and method of protection. The
surge protector requires three outlets and a minimum rating of 1500 watts. Either piece of
equipment should be listed by Underwriters Laboratory (UL) or a similar listing body, certified
by Canadian Standards Association (CSA), and marked Conformité Europeénne (CE) for
nondomestic use.

For research use only. Not for use in diagnostic procedures.

Technical Overview

21

Advertising