3.2.8 Static Pressure Setup

The differential pressure transmitter can display the

static pressure also.

(1) Setting of the unit for static pressure
Follow the procedure below to change the static

pressure unit.

• Procedure to call up the display

DD and DTM

(excluding EJX_

HART 5[1.2])

[Root Menu] → Basic setup →

SP setup →

EJX_HART 5[1.2]

DTM

Configuration → Process Input (or

Static Pressure Sensor) →

→ SP Unit

Select the unit for static pressure

(Refer to subsection 3.2.2 Unit)

(2) Setting of the measuring range for static

pressure

Follow the procedure below to change the lower

range value (LRV) and upper range value (URV).

• Procedure to call up the display

DD and DTM

(excluding EJX_

HART 5[1.2])

[Root Menu] → Basic setup →

SP setup → SP Range → Keypad

input →

EJX_HART 5[1.2]

DTM

Configuration → Static Pressure

Sensor →

→ SP LRV

Set the lower range value (0 %) of

static pressure

→ SP URV

Set the upper range value (100 %)

of static pressure

(3) Selection of Gauge pressure and Absolute

pressure

Either the gauge pressure or absolute pressure can

be selected to display on the LCD display.
Absolute pressure is selected when the transmitter

is shipped.

• Procedure to call up the display

DD and DTM

(excluding EJX_

HART 5[1.2])

[Root Menu] → Basic setup →

SP setup → SP A/G Setup →

EJX_HART 5[1.2]

DTM

Configuration → Static Pressure

Sensor →

→ A/G Select

Select “Gauge” or “Absolute”

→ Atm. Pres Value 0.1013 MPa when the transmitter

is shipped

(4) Selection of pressure side
Either the high or low pressure side of capsule can

be selected to monitor the static pressure.
High pressure side is selected when the transmitter

is shipped.

• Procedure to call up the display

DD and DTM

(excluding EJX_

HART 5[1.2])

[Root Menu] → Basic setup → SP

setup →

EJX_HART 5[1.2]

DTM

Configuration → Static Pressure

Sensor →

→ SP H/L Select

Select “High” or “Low”

3.3

Detailed Setup

3.3.1 Bi-directional Flow Measurement

(a)

Bi-dir mode enables selection of 50% output at

an input of 0 mmH

2

O.

• Procedure to call up the display

DD and DTM

(excluding EJX_

HART 5[1.2])

[Root Menu] → Detailed setup →

Signal condition →

EJX_HART 5[1.2]

DTM

Configuration → Analog output →

→ Bi-dir mode

Select “On” or “Off”

(b) Combining

Bi-dir mode with Xfer fnctn

provides a square root output computed

independently for 0% to 50% output and for

50% to 100% output.

20 mA (100% display)

4 mA (−100% display)

Output mode “LINEAR”

LRV

HRV

F0304.ai

20 mA (100% display)

Low Cut

4 mA (−100% display)

Output mode “SQUARE ROOT”

LRV

HRV

Advertising
This manual is related to the following products: