3 alarm setting procedure, Default values -15, Caution – Yokogawa Integral Oxygen Analyzer ZR202 User Manual
Page 108
IM 11M12A01-05E
8-11
8. Detailed Data Setting
8.3.3
Alarm Setting Procedure
Set the alarm setpoints following Table 8.6 listing parameter codes.
Table 8.6 Alarm Parameter Codes
T8.3.3.1.EPS
Set item
Parameter code
Set value
Oxygen concentration high-high alarm setpoint
D01
0-100% O
2
Oxygen concentration high-alarm setpoint
D02
0-100% O
2
Oxygen concentration low-alarm setpoint
D03
0-100% O
2
Oxygen concentration low-low-alarm setpoint
D04
0-100% O
2
Humidity high-high alarm setpoint
D05
0-100% H
2
O
Humidity high-alarm setpoint
D06
0-100% H
2
O
Humidity low-alarm setpoint
D07
0-100% H
2
O
Humidity low-low-alarm setpoint
D08
0-100% H
2
O
Mixing ratio high-high alarm setpoint
D11
0-1 kg/kg
Mixing ratio high-alarm setpoint
D12
0-1 kg/kg
Mixing ratio low-alarm setpoint
D13
0-1 kg/kg
Mixing ratio low-low-alarm setpoint
D14
0-1 kg/kg
Oxygen concentration alarm hysteresis
D30
0-9.9% O
2
Humidity hysteresis
D31
0-9.9% H
2
O
Mixing ratio hysteresis
D32
0-0.1 kg/kg
Delay alarm action
D33
0-255 seconds
Oxygen concentration high-high alarm detection
D41
0: Not detected
1: Detected
Oxygen concentration high-alarm detection
D42
0: Not detected
1: Detected
Oxygen concentration low-alarm detection
D43
0: Not detected
1: Detected
Oxygen concentration low-low-alarm detection
D44
0: Not detected
1: Detected
Humidity high-high alarm detection
D45
0: Not detected
1: Detected
Humidity high-alarm detection
D46
0: Not detected
1: Detected
Humidity low-alarm detection
D47
0: Not detected
1: Detected
Humidity low-low-alarm detection
D48
0: Not detected
1: Detected
Mixing ratio high-high alarm detection
D51
0: Not detected
1: Detected
Mixing ratio high-alarm detection
D52
0: Not detected
1: Detected
Mixing ratio low-alarm detection
D53
0: Note detected
1: Detected
Mixing ratio low-low-alarm detection
D54
0: Not detected
1: Detected
CAUTION
Even with alarms set, if “Not detected” has been set in the above alarm detection, no
alarm is issued. Be sure to set “Detected” in the above alarm detection if you use alarm
features.