5 wiring for contact input, 1 cable specifications, 2 wiring procedure – Yokogawa Integral Oxygen Analyzer ZR202 User Manual
Page 70: Zr202g high-temperature humidity analyzer -1
IM 11M12A01-05E
5-7
5. Wiring
5.5 Wiring for Contact Input
The converter can execute specified function when receiving contact signals.
To use these contact signals, proceed wiring as follows:
Converter
Terminal box
DI-1
DI-C
DI-2
Contact input 1
Contact input 2
F5.7E.EPS
Figure 5.14 Contact Input Wiring
5.5.1
Cable Specifications
Use a 2-wire or 3-wire cable for this wiring. Depending on the number of input(s),
determine which cable to use.
5.5.2
Wiring Procedure
(1) M4 screws are used for the terminals of the converter. Each wire in the cable should
be terminated in the corresponding crimp-on terminal.
(2) The ON/OFF level of this contact input is identified by the resistance. Connect a
contact input that satisfies the descriptions in Table 5.2.
Table 5.2 Identification of Contact Input ON/OFF
T5.2E.EPS
Closed
Open
Resistance
200 V or less
100 kV or more