6 setting display item – Yokogawa Integral Oxygen Analyzer ZR202 User Manual

Page 84

Advertising
background image

IM 11M12A01-05E

7-11

7. Startup

7.6 Setting Display Item

Display items are those items that are displayed on the basic panel display. Parameter
code A00 or F08 is used to set the display items as shown in the table below. If the
humidity analyzer /HS option was specified at the time of purchase, the equipment is a
humidity analyzer. For other than the above, the equipment is set to oxygen concentra-
tion at the factory before shipment. If mix ratio is to be measured, change the existing
setting as follows.

Additionally, when humidity analyzer is selected in the Detector Type Setting in the
previous section, the display item will be humidity if data initialization is performed.

Table 7.4 Display Items

T7.4.EPS

Values set with

Items displayed on the basic panel display

A00 or F08
0

Indicates the oxygen concentration.

1

Indicates the humidity.

2

Indicates the mix ratio.

3

Displays an item for the current output. If the output damping

has been set for the current output, values involving the output

damping are displayed.

Table 7.5 Display Item Setting Procedure

T7601.EPS

Display after the password has been entered.

Touch the [

] key to move the position of the digit that is

flashing to the right.

Touch the [

] key to change the parameter code to 0. The

number will change from 1 to 2 to 3 9 and back to 0.

Touch the [ENT] key to display the current set value.

Set the mix ratio. Touch the [ENT] key to change the

number to 2.

Touch the [ENT] key. The number 2 will be flashing.

Touch the [ENT] key again to stop flashing.

Touch the [ENT] key to return to the parameter code

selection display.

Touch the [

] and [ENT] keys simultaneously to return to

the basic panel display on which the mix ratio appears.

.

.

.

.

.

.

.

.

.

.

.

.

.

.

.

.

.

.

.

.

.

The symbol ( ) indicates that the corresponding keys are being touched, and the light

characters indicate flashing.

A

01

A0

1

A0

0

0

2

2

2

A

00

0.000

Switch operation

Display

Description

ENT

ENT

ENT

ENT

ENT

ENT

ENT

ENT

ENT

Advertising