5 setting the computation channels (/m1 option) – Yokogawa µR20000 User Manual

Page 27

Advertising
background image

2-14

IM 04P01B01-61E

2.5 Setting the Computation Channels (/M1 Option)

Click the

Math tab. You can also choose Math Channels from the Setting menu.

Tag

Alarm delay time

Channel recording color (dot model)

Turning ON/OFF computation

Equation

Constants used

in equations

Click twice to unselect the channel.

Double-click to open a setup dialog box for the respective channel.

Drag the pointer while holding down the left mouse button to select multiple channels.

Recording span

Rightmost

value

Unit

Leftmost

value

Decimal point

position

Alarm 1

Output relay

number

Alarm value

Alarm type

Sets the settings of the first channel in the selected

range to the other channels.

Partial expanded recording

Bar graph display mode

Turns ON/OFF periodic printout

Types of measured values

to be printed periodically

Turn trend recording ON/OFF (dot model)

Recording zone

Alarm 4

Sets the default value.

Toggles ON/OFF collectively for each click.

Sets the minimum possible value.

Sets the maximum possible value.

The tool buttons apply to the selected range of channels when channels are selected.

They apply to all channels when a channel is not selected.