3 setting the moving average (dot model), 3 setting the moving average (dot model) -3, 3 settng the movng average (dot model) – Yokogawa µR10000 User Manual

Page 117: App index

Advertising
background image

6-3

IM 04P01B01-01E

Setup

Operat

ons
for

Conven

ent
Funct

ons
(Sett

ng

Mode)

1

2

3

4

5

6

7

8

9

10

11

12

App

Index

6.3 Settng the Movng Average (Dot Model)

This section explains the details of setting the moving average function on measurement
channels on the dot model.

Procedure

1.

Hold down the

MENU

key for 3 seconds to enter Setting mode.

2.

Press the

key or

SHIFT +

key to select

Aux and then press the key.

3.

Press the

key or

SHIFT +

key to select

Movng_AVE and then press the

key.

4.

Set each item and press the key.

Use the

key or

SHIFT +

key to select values.

For the procedure on how to enter values or characters, see section 4.2.
If you press the

ESC key in the middle of the operation, the settings entered up to

then are cancelled, and the display returns to a higher level menu.

5.

When the

Settng complete screen appears, do either of the following:

Press the key to set other channels.
If you are done, press the

ESC key.

6.

Hold down the

MENU

key for 3 seconds to return to Operation mode.

Set the channel range.

Set the number of samples of the

moving average. Moving average is

not performed when Off is selected.

The new setting takes effect.

ESC/?

First channel Last channel

Set=Aux

Aux=Moving_AVE

CH=01-01

No.of samples=Off

01-01 Channel

Setting complete

Explanaton

The Number of Samples of Movng Average

The number of samples can be set to a integer between 2 and 16. Select Off when not
using the moving average function.

<Related Topcs> Enabling the moving average function: Section 7.11

Advertising