Electronic module
(see detail „A”)

See table 7 for details
of this connection

Disconnect
this end when
installing.

See table 5
for dressing of
cable at this end.

Outside braid makes contact to
housing. Braid is compressed for a
good electrical connection.
See table 6 and 7.

1/2 inch conduit may be connected
directly to connectors via 3-piece
union/coupler.

Cable must be pushed into stainless steel
fittings when installing compression nuts
to ensure that the braid is properly seated
for a good electrical connection (both ends).
See tables 6 and 7.

See table 5 for
dressing of cable
at this end.

VFM 3100 body.

External earth (ground)
connection screw –
cenelec flameproof only.

Pre-assembled and dressed
junction box

Braided shield and drain wire must
be in contact at this end of cable.
Note: do not disassemble to install.

Remote
cable

Output
signal
wires

brown

red

orange

yellow

Detail “A”

Figure 9. Overall View of Remote Mounted VFM 3100