ELECTRICAL CONNECTIONS

30

MFC 300

www.krohne.com

02/2012 - 4000498602 - MA MFC 300 R03 en

4.4.2 Connection of signal cable, wall-mounted housing

1 Open the housing cover.
2 Open the cover and guide the prepared signal cable through the cable entry.
3 Connect the twisted shielding to terminal S.
4 Connect the electrical conductors to terminals +, -, A, B.
5 Press the plug into the connector.
6 Close the cover and the housing cover.

Figure 4-2: Electrical connection of signal cable, wall-mounted housing

a = white
b = green
c = black
d = red

INFORMATION!
Each time a housing cover is opened, the thread should be cleaned and greased. Use only resin-
free and acid-free grease.
Ensure that the housing gasket is properly fitted, clean and undamaged.