Operation – KROHNE MFC 300 EN User Manual

Page 75

Advertising
background image

OPERATION

6

75

MFC 300

www.krohne.com

02/2012 - 4000498602 - MA MFC 300 R03 en

B) Manual calibration

Under certain conditions, zero calibration is not possible and will be cancelled:

• The product is still flowing. The shut-off valves are not closed tightly enough.
• There are still gas bubbles in the fluid.

Remedy: Flush sensor and repeat calibration

With some media, it may be difficult to perform zero calibration. In such cases, there are various
methods to still achieve good zero calibration:

Key

Display

Description and setting

>

A

Quick Setup

Press and hold for 2.5 s, then release

the key.

2 x ↓

C

Setup

3 x >

C1.1.1

Zero Calibration

>

Calibrate Zero?

Break

3 x ↓

Calibrate Zero?

Manual

Zero Calibration

+XX.XXX%

Display currently stored zero

calibration in %.

(Warning, value can be changed!)

Possible manual entry of zero

calibration.

Storing the displayed zero

calibration.

5 x ^

Save Configuration?

Yes

^

Display page

Medium

Possible solutions

Media that tend to evaporate or outgas

Increase pressure.

Two-phase media (slurries), that contain solids that

may fall out.

Only fill the sensor with the carrier medium.

Two-phase media in which the solids or gas-

forming components cannot be separated.

Fill sensor with another liquid, e.g. water.