Commissioning / operation – KROHNE UFC 400 PROFIBUS PA-PROFIBUS DP User Manual

Page 34

Advertising
background image

4

COMMISSIONING / OPERATION

34

UFC 400

www.krohne.com

05/2014 - 4003118502 - AD UFC 400 PROFIBUS R02 en

76

3

("S-") Proc: Signal lost

(default classification / 9)

Process related fault(s) detected concerning the "sensor". E.g. an

unknown accuracy. The measurement accuracy may be reduced.

But still the measurement value follows the source. Check

fault(s) detected / 4 - if the fault(s) cannot be cleared contact the

service department of the manufacturer with an indication of the

problem(s).

77

3

("S-") Proc: Empty Pipe

(default classification / 9)

Process related error(s) detected concerning the "sensor". E.g.

no liquid present. Related output values are invalid. A repair or

replacement of the device can be necessary. Check error(s)

detected / 4 - restart the device – if the error still occur contact

the service department of the manufacturer.

78

3

("S-") Electr: IO Connection

(default classification / 9)

Converter related fault(s) detected concerning interconnection of

modules and terminals. E.g. an open or short circuit at the related

terminal. The measurement accuracy may be reduced. But still

the measurement value follows the source. Check fault(s)

detected / 4 - if the fault(s) cannot be cleared contact the service

department of the manufacturer with an indication of the

problem(s).

81

3

("I-") Electr: Info. Operation

(classification fixed - not changeable)

Converter Electronic related information. Just information. No

impact on measurement expected. Check information just

present / 7.

82

3

("I-") Config: No Meas. Value

(classification fixed - not changeable)

Configuration / measurement value related information. Just

information. No impact on measurement expected. Check

information just present / 7.

99

3

F - No Meas. Value

(classification fixed - not changeable)

Measurement value not available (or no more available) / 8.

(status bit "F – Electronics" will be set too)

100

3

F - Parameter Update Error

(classification fixed - not changeable)

Error occurred during parameter update of PROFIBUS

parameters / 8.

(status bit "F – Configuration" will be set too)

101

3

F - No Device Communication

(classification fixed - not changeable)

No communication between basic device and PROFIBUS module /

8.

(status bit "F – Electronics" will be set too)

Bit Number

sup

p

orted

De

sc

ription

Remedy