Vaisala GMP343 User Manual

Page 68

Advertising
background image

User’s Guide ______________________________________________________________________

66 ___________________________________________________________________M210514EN-E

3.

The GMP343 is now measuring the second reference gas (1007 ppm).
Use command R and let the reading stabilize for several minutes.
Calculate the average reading (reading 2).

>

>r

1067.1

1066.8

1067.2

1066.7

1066.6

4.

Turn off the calibration mode with command CALIB OFF.

>calib off

Calibration mode stopped.

5.

Next feed the readings: use command LCI to feed the first and second
reference gas readings (28 ppm and 1066 ppm), and the corrected
values (= the concentrations of the reference gases, 0 ppm and 1007
ppm).

>lci

Reading Reference

0.00 0.000

1000.00 1000.000

NOTE: Entering new correction

values all previous correction points!

Abort without losing correction points

by using ESC.

1. Reading ? 28

Reference ? 0

2. Reading? 1066

Reference? 1007

6.

Confirm the adjustment with command LC ON.

>lc

LC : OFF ? on

7.

Check the reading with the second reference gas, 1007 ppm.

>

r

1005.4

1006.2

1007.1

1007.1