Error states, Table 7, Troubleshooting errors – Vaisala GMP343 User Manual

Page 74

Advertising
background image

User’s Guide ______________________________________________________________________

72 ___________________________________________________________________M210514EN-E

Error States

In case of an error function the outputs are as follows:

-

Serial line: In STOP mode the error code and description (see the
Table 6 below) is shown when keying in the command: ERRS. In
POLL or RUN mode the error flag is set to 1 (if the error flag is
included in the message format, see FORM-command). Reset the
error state with a command ERRS R.

-

Analog output shows the error level (can be set by using the AERR-
command. Factory setting: 5 V or 20 mA).

Table 7

Troubleshooting Errors

Error code

Error description

Interpretation

Action

E01

EEPROM checksum failure Internal transmitter failure

Return the device to
Vaisala Service

E02

IR source failure

Internal transmitter failure

Return the device to
Vaisala Service.

E03

FPI failure

Internal transmitter failure

Return the device to
Vaisala Service.

E04, E05

Heater failure

Internal transmitter failure

Return the device to
Vaisala Service.

E06

Temperature measurement
failure

Operation temperature is
out of allowed range.
Analog output: Error level
is shown if temperature
compensation is enabled.
Otherwise, output is
normal.

Ensure that the operating
temperature is -45...+85
°C (-49...185 °F). In case
of constant error, return
the device to Vaisala
Service.

E07

Measurement signal level
too low

Measurement chamber is
contamined or the lamp is
degraded.

Change the filter and clean
the optics according to the
instructions in section

Changing the Filter and
Cleaning the Optics
(Diffusion Model Only) on
page 70

. In case of

constant error, return the
device to Vaisala Service.

W01

Watchdog reset occured

Software defect

In case this warning
appears frequently, return
the device to Vaisala
Service.

W02

Stack overflow

Software defect

In case this warning
appears frequently, return
the device to Vaisala
Service.

Advertising