General specifications, General specifications -11 – Agilent Technologies B2201A User Manual

Page 27

Advertising
background image

Agilent B2200 User’s Guide, Edition 2

1-11

Introduction

Specifications

General Specifications

Temperature

range:

Operating: 5

°C to 35 °C

Storage:

− 20 °C to 70 °C

Humidity range:

Operating: 5 % to 70 % R.H., non-condensing
Storage:
< 80 % R.H. at 35

°C, < 60 % R.H. at 65 °C, non-condensing (B2200A/B2210A)

< 80 % R.H. at 65

°C, non-condensing (B2201A/B2211A)

Altitude:

Operating: 0 to 2,000 m (6,500 ft)
Storage: 0 to 15,240 m (50,000 ft)

Regulatory

compliance:

Safety: CSA C22.2 No. 1010.1/IEC 1010-1

EMC: CISPR 11 Group 1 class A&EN50082-1

Power

requirement:

90 to 264 V (continuous), 47 to 63 Hz, 2 A/200 VA maximum

Number of slots:

4 slots for 48 mm height switch module

Dimensions:

B2200A/B2201A: 430 mm W

× 320 mm H × 600 mm D

B2210A/B2211A: 395 mm W

× 48 mm H × 500 mm D

Weight:

B2200A/B2201A: approximately 14.0 kg
B2210A: approximately 5.0 kg
B2211A: approximately 3.5 kg

Number of ports:

SMU input: 8 triaxial ports (with Guard)

AUX input: 6 coaxial ports (two ports for capacitance measurement)
Output: 12 triaxial ports (with Guard), maximum 48 ports