Agilent Technologies B2201A User Manual
Page 48
2-14
Agilent B2200 User’s Guide, Edition 2
Installation
Output Connections
Table 2-4
Connection to the DUT Interface
CAUTION
Never connect the guard terminal to any output, including circuit common, chassis
ground, or the guard terminal of any other unit. Doing so may result in an
emergency condition.
For highly accurate current forcing and measurements while minimizing leakage,
surround all force and sense lines from SMU by a guard as far as possible, and make
cables stable by taping.
For reducing capacitance measurement error, tape the cables to any grounded
materials such as shielding box. If you use probe card, grounding the probe card also
reduces capacitance measurement error.
Kelvin connections
non-Kelvin connections
This connection is available only for the
Kelvin connectors.
This connection can cancel effects of cable
resistance by connect sense line and force line
as close as possible to DUT terminal.
Following connection is for the Kelvin
connectors. For the triaxial connectors,
ignore SENSE terminal, and make
connection only for FORCE terminal.
Measurement data will include residual
resistance from the connection cable.
to
DUT
Coaxial cable
Triaxial
connector
Insulator
Plate
Guard
Sense
Common
Guard
Force
Common
to
DUT
Coaxial cable
Triaxial
connector
Insulator
Plate
Guard
Sense
Common
Wire
Guard
Force
Common