39

For adjusting the zero, operate the gauge under the same con-
stant ambient conditions and in the same mounting orientation
as normally.

If you are using a seal with centering ring and filter, check

that they are clean or replace them if necessary
(

→ "Deinstallation").

Put the gauge into operation and operate it at atmospheric

pressure for at least 10 minutes.

Press the <ADJ> button with a pin (max. ø1.1 mm) and the

ATM adjustment is carried out: The Pirani sensor is ad-
justed to 1000 mbar (duration

≈5 s).

ATM

HV

If the pressure value 1000 mbar is output at the

measurement value output, the adjustment has been
successful. Otherwise, repeat the adjustment proce-
dure.

Evacuate the vacuum system to p < 10

-5

mbar and wait at

least 2 minutes.