Pr eliminar y, Page 15 – Practical Instrument Electronics 850 User Manual

Page 17

Advertising
background image

Page 15

Pr

eliminar

y

To

Sensor

Typical

2-Wire

Transmitter

(Disconnected)

+ IN - REF +OUT-

Receiver

Powers External

2-Wire Transmitter

Power Supply

(2 to 60 VDC)

LO

4.000

mA

2-WIRE

5 6

+ -

Connect the output leads of the PIE 850 to the inputs of the
device being calibrated, making sure to check polarity. Red lead
to the plus (+) input and black lead to the minus (-) input.

Instantly output your SPAN and ZERO output settings by moving
the EZ-CHECK switch between HI and LO (defaults to 20 & 4 or
50 & 10 mA). You may also select any third output setting (such
as mid-range) using the SET position on the EZ-CHECK switch.
The output is adjusted in 0.001 mA (0.01%) increments by
turning the knob

e

. Press and turn the knob for faster dialing

with 0.100 mA (1.00%) increments.

2 Wire SIM mA, 2 Wire SIM % (Percent of 4-20 or 10-50 mA)

T/C

All mA∙ Read V

-

2

3

RTD

RTD

6

5

4

+

Source & Read Hz∙ mV

Source V∙ pH

-

+

1

-

All milliamp connections

Note: With LOOPSCOPE on the 850 occasionally performs a
test on the loop. During this short test the loop mA signal will be
affected.