Pr eliminar y – Practical Instrument Electronics 850 User Manual

Page 24

Advertising
background image

Page 22

Pr

eliminar

y

Voltage Output Signal

Controller

Transmitter

Power Supply

MAX

8.300

mV

2 1

+

-

Read mV/V

Choose this function to measure from -99.999 to 99.999
millivolts, -999.99 to +999.99 mV, 0.000 to 10.250 V dc or 0.00
to 60.00 V dc.

Move the power switch

w

to READ then Double Click the

EZ-DIAL knob to get into the Menu. Turn the knob to scroll
through the settings and press the knob to make your selection.
Select V for the FUNCTION and 1V, 10V, 60V or 100 mV for the
RANGE.

Connect the red input lead (+) of the PIE 850 to the more
positive point of the break and the black input to the more
negative point.

Signals above the maximum scale are limited by protection
circuitry with “OVER RANGE” flashed on the display and the red
OVERLOAD LED lit.

The PIE 850 measures the input signal and constantly updates
the display with the current reading. Move the EZ-CHECK switch

q

to MAX to see the highest reading and to MIN to see the

lowest reading. Press and hold the knob

e

to clear the MAX and

MIN readings.

T/C

All mA∙ Read V

-

2

3

RTD

RTD

6

5

4

+

Source & Read Hz∙ mV

Source V∙ pH

-

+

1

-

Connections for
Read mV ranges

Connections for Read
10 V & 60V ranges

Advertising