Publication GMSI00-UM001A-EN-E - February 2005

Enpac Ex Parameters 115

Ext Trig Slope

The input signal slope to be used for externally
triggering the measurement.

+ve - Defines the trigger event to occur on a rising

edge.

-ve - Defines the trigger event to occur on a falling

edge.

Options: +ve

-ve

Level

The amplitude level for externally triggering the
measurement. The level is a percentage of the
trigger full scale range.

Bin Zeroing

Enables/disables bin zeroing. When set to "Off," the
bins below the selected high pass filter are not
zeroed. That is, data still appears but is attenuated
by the filter shape.

Set bin zeroing to "On" when you want to be sure
unwanted frequencies do not affect data. Set bin
zeroing to "Off" when you want to be aware of the
presence of unwanted frequencies.

Options: On

Off

Offroute User Parameters

Parameter Name

Description

Values/Comments