Fault Location Mode

75

Using Markers with Recalled Traces
When a trace is recalled, the screen displays both the current trace and

the recalled trace and there is an additional marker option, single / dual

mode. In single mode, the markers function normally; the site analyzer

displays the frequency and trace magnitude at each marker. In dual

mode, the markers show the actual magnitude of both traces and the

difference in magnitude between the recalled trace and the current

reading. All markers follow the current trace only, and only three mark-

ers are available in dual mode, M1, M2, and M3. When you select dual

mode, the markers move as described in Move Markers, page 72.
1. Be sure that you are in either Measure Match mode or Fault

Location mode.

2. Recall a trace (refer to Recalling a Saved Trace, page 92).
3. Follow the numbered steps in Figure 37 on page 75.
Figure 37 Using Markers with Recalled Traces

Step

Action

1

Press the Marker key

2

Press the Mode softkey to select single
mode or dual mode

3

Select a marker to move

4

Press the left or right scroll keys to move
the marker

5

Press the Esc key to return to the test screen