4 working with connected balance, 5 karl fischer titrations – Metrohm TiNet 2.5 User Manual

Page 49

Advertising
background image

TiNet 2.5 Compliant

Titrations 47

5.4.4

Working with connected balance

Weighing data of a connected balance can be transferred.
The receipt of weighing data is confirmed by the PC with a beep.
Important: Depending on the work load of your PC, you should wait a few
seconds between transmissions of the individual weighing data.
The balance is always assigned to the active parallel titration. If the

Silo

option is disabled, the weighing data are sent directly to the current sam-
ple data row. If the

Silo

option is enabled, they are sent to the silo.






5.4.5

Karl Fischer titrations

Start/Stop

Start conditioning with button

<Conditioning>

.

Wait for a stable drift and start the titration. The titra-
tion can only be started if "conditioning is ok". The
titration start stops the conditioning; it is therefore
important that there is not a great delay between the
start and the beginning of the titration. After the PC
method is terminated, the titration vessel is again
conditioned.

This key stops the titration as well as the condition-
ing.


Prerequisites in the method
Conditioning has to be activated in the titration blocks of the method.
Conditioning is possible for the following blocks:

• KFC (conditioning is always active)
• KFT
• SET
• Titrino method (conditioning is active in the Titrino method)