4 restrict method selection, 4 results, 1 sign results – Metrohm TiNet 2.5 User Manual
Page 68
66 • Working according to 21 CFR 11
TiNet 2.5 Compliant
5. If desired, print a method report including all method data stored in
the method database by selecting the
Method report
menu item using
the right mouse key.
6. Close
the
Method administration
window.
7.3.4
Restrict method selection
1. Enable
the
Extra/Methods/Use signed methods only
option to make sure
that only signed database methods can be selected in the
Method
field of the
Sample data
tab and for the silo.
2. To make sure that users cannot disable the
Use signed methods only
option, disable this menu item for the user group under
User admini-
stration/'Group'/Access policies/Menu items
(see section 3.2.2).
7.4 Results
7.4.1 Sign
results
1. Select
Extra/Results/Results overview
to open the
Results overview
win-
dow (see section 6.3).
2. Select the desired result, which should be signed.
3. Using the right mouse key, select the
Sign-off result
menu item. The
Sign-off
window appears.
4. Enter your user name and password and select or enter a sign com-
ment. Click on
<Sign-off>
to sign the result. Afterwards, click on
<Sign-off>
.
5. If desired, print a result report including all determination data stored
in the result database by selecting the
Result report
menu item using
the right mouse key.
6. Close
the
Results overview
window.