4 restrict method selection, 4 results, 1 sign results – Metrohm TiNet 2.5 User Manual

Page 68

Advertising
background image

66 Working according to 21 CFR 11

TiNet 2.5 Compliant

5. If desired, print a method report including all method data stored in

the method database by selecting the

Method report

menu item using

the right mouse key.

6. Close

the

Method administration

window.

7.3.4

Restrict method selection

1. Enable

the

Extra/Methods/Use signed methods only

option to make sure

that only signed database methods can be selected in the

Method

field of the

Sample data

tab and for the silo.

2. To make sure that users cannot disable the

Use signed methods only

option, disable this menu item for the user group under

User admini-

stration/'Group'/Access policies/Menu items

(see section 3.2.2).

7.4 Results

7.4.1 Sign

results

1. Select

Extra/Results/Results overview

to open the

Results overview

win-

dow (see section 6.3).

2. Select the desired result, which should be signed.
3. Using the right mouse key, select the

Sign-off result

menu item. The

Sign-off

window appears.

4. Enter your user name and password and select or enter a sign com-

ment. Click on

<Sign-off>

to sign the result. Afterwards, click on

<Sign-off>

.

5. If desired, print a result report including all determination data stored

in the result database by selecting the

Result report

menu item using

the right mouse key.

6. Close

the

Results overview

window.