Select the requested standard from the set.
Labels on the back identify each standard.

Note the mini-CD that contains the
“Standards File”. This file is used only during
IPV Setup.

7.2

Evaluation Diagnostics

Evaluation Diagnostics are used to verify that the instrument is operating within allowable
parameters. These tests should be run approximately once per week.

This information is meant to guide the user through the tests in an expeditious manner. A more
complete description of these tests is given in the Vision Manual, in the Diagnostics section. A
discussion of the theory and interpretation of results is provided in the Vision manual.

7.2.1

Performance Test

Performance Test is a comprehensive test of instrumental performance, and is the final assurance that
the instrument is ready to run samples. The key items verified during this test are:

Instrument Noise in each of four wavelength regions

Internal Wavelength Performance (wavelength positions on non-traceable, internal reference

materials.)

Internal Wavelength Precision (Repeatability)

NIR Gain

Visible Gain

The test is initiated as follows:

Select Performance Test from the
Diagnostics menu. Click on Run
Performance Test.

As the test runs, a screen like that shown below is displayed. When finished, a message box is
displayed to indicate test completion and status.