4.1. Error and special messages

794 Basic Titrino

136

E44

The RS interface parameters are no longer the same

for both instruments.

E45

The receive buffer of the Titrino contains an

incomplete string (missing L

F

). Transmission of the

Titrino is thus blocked.

Exit: Send

L

F

or <QUIT>.