2.7. Parameters, key <PARAM>

794 Basic Titrino

40


>preselections

conditioning: OFF





display drift: ON

cond.


req.ident: OFF

cond.





req.smpl size: OFF

cond.



activate pulse: OFF

cond.

Preselections for the sequence

Automatic conditioning of titration vessel. (ON, OFF)

If conditioning is "on", between the titrations the titration

solution is constantly maintained at the (1st) end point.

When conditioning is performed, the volume drift can

be displayed during the conditioning:

Display of drift during conditioning (ON, OFF).

Volume drift.

Request of identifications after start of titration (id1,

id1&2, all, OFF)

After start, sample identifications can be requested

automatically: Only id1, id1 & id2, all three id's or no

inquiries.

Request of sample size after start of titration (value, unit,

all, OFF)

"all" the value and the unit will be requested.

Pulse output on I/O line 6 (L6, pin 1) of the remote

socket (first, all, cond., OFF)

see page 163.