2.7. Parameters, key <PARAM>

794 Basic Titrino

30






low lim.1 pH -20.00

up lim.1 pH 20.00













fix EP1 at pH OFF











pK/HNP: OFF

"window": Only EP's in specified windows are

recognized.

"OFF":

Evaluation switched off.

If "window" is selected, lower and upper limits of

windows are inquired.

Lower limit of window 1

and

upper limit of window 1

(Input ranges for both inquiries depend on the

measured quantity):

pH: 0...±20.00,

OFF

U, Ipol: 0...±2000 mV, OFF

Upol:

0...±200.0 µA, OFF)

<CLEAR> sets "OFF".

Only equivalence points are recognized which lie within

the set lower and upper limits. The equivalence point

numbering is defined with the windows, see page 34.

Window inquiries continue until the lower limit is set to

"OFF". Up to 9 possible windows.

Always set both limits to ≠ OFF for a valid window.

Interpolation of volumes at fixed times (input range

depends on the measured quantity:

pH: 0...±20.00,

OFF

U, Ipol: 0...±2000 mV, OFF

Upol:

0...±200.0 µA, OFF)

<CLEAR> sets "OFF".

If a fix end point has been set, the volume value for the

input measured value is interpolated from the curve,

see also page 35. The volume values are available as

C5X.

Fix EP's are inquired until "OFF" is set. Up to 9 fix EP's.

Evaluation of pK or HNP (ON, OFF)

pK evaluation in case of pH titrations and half

neutralization potential for U, see page 35.


>preselections

req.ident: OFF





req.smpl size: OFF



activate pulse: OFF

Preselections for the sequence

Request of identifications after start of titration. (id1,

id1&2, all, OFF)

After start, sample identifications can be inquired

automatically: Only id1, id1 & id2, all three id's or no

inquiries.

Request of sample size after start of titration (value, unit,

all, OFF)

"all" requests the value, then the unit.

Pulse output on I/O line "activate" (L6, pin 1) of the

remote socket (ON, OFF)

see page 163.