3 mains connection – Metrohm 763 PVC Thermomat Manual User Manual

Page 20

Advertising
background image

2.3 Mains connection

763 PVC Thermomat

13

Connect one end of the 6.1816.010 Silicone tubing to con-

nection piece 22

22.

Plug the other end of the 6.1816.010 Silicone tubing into the

corresponding exhaust collection tube opening.

Close the unused openings on the exhaust collection tube

with the included E.400.0010 stoppers.

3 Connecting the exhaust collection tube

Attach a piece of suitable tubing to the exhaust collection

tube connection and connect it to an active aspirating device

(e.g. water-jet pump).

2.3

Mains connection

Follow the instructions below for connecting to the power supply. If

the instrument is operated with a mains voltage set wrongly and/or

wrong mains fuse, there is a danger of fire!

2.3.1

Checking the mains voltage

Before switching on the 763 PVC Thermomat for the first time, check

that the mains voltage indicated on the model plate 15

15 (see Fig. 3)

matches the local mains voltage. If this is not the case, call the

Metrohm service.

2.3.2

Exchange of fuses

Two fuses 4 A/slow-blow for 115 V or 2 A/slow-blow for 230 V are in-

stalled in fuse holder 14

14 of the 763 PVC Thermomat as standard.

Ensure that the instrument is never put into operation with fuses of

another type, otherwise there is danger of fire!

For changing fuses, proceed as follows:

1 Disconnect mains cable

Disconnect mains cable from mains connection plug 13

13 of the

763 PVC Thermomat.

2 Remove fuse holder

Using a screwdriver, loosen fuse holder 14

14 above the mains

connection plug 13

13 and take out completely.