Page 60
4.4 Calibration functions
763 PVC Thermomat
53
If the temperature defined in the method has not been reached then the
heating is started automatically. The message
Waiting until temperature is
reached...
appears in the window.
When the temperature defined in the method has been reached the
message
Measuring temperature. Please wait...
appears in the window.
The three following values are shown below this message:
Block temperature
The temperature in the heating block measured
by the internal temperature sensor.
Temperature ext. sensor
The temperature in the reaction vessel meas-
ured by the external temperature sensor.
Delta T
The temperature correction determined from
the difference between
Block temperature – Tem-
perature ext. sensor
.
As soon as both the measured temperatures are stable the message
Measured temperature is now stable!
appears. The determined tempera-
ture correction
Delta T
can now be included in the method with
<Accept>
.
The message
Delta T is now set in the corresponding method
then appears.