7 precolumns and separating columns, 1 precolumns, Precolumns and separating columns – Metrohm 861 Advanced Compact IC User Manual

Page 40: Precolumns

Advertising
background image

2 Installation

861 Advanced Compact IC / Instructions for Use 8.861.1033

30

sure pump.

Close the rotary knob on purge valve 37 by turning it in a

clockwise direction (see Figure 3 and Figure 4 and Figure 5).

Remove the syringe from connection 2.

5 Rinse pulsation dampener

Place a beaker beneath the column connection capillary 26.

Click the

<On>

button for

IC pump

to switch on the high-pres-

sure pump and rinse the pulsation dampener 60 filled with

isopropanol for approx. 10 min with eluent.

Click the

<Off>

button for

IC pump

to switch off the high-pres-

sure pump.

6 Reduce flow rate

Reset the original flow rate under

Flow

(e.g.

0.5 mL/min

).

Click

<Send to unit>

to send this value to the 861 Advanced

Compact IC.

Click

<Save>

to save the values.

2.7

Precolumns and separating columns

2.7.1 Precolumns

The use of easily exchangeable precolumns protects the separating

columns and prolongs their lifetime. The precolumns available from

Metrohm are either real precolumns or precolumn cartridges, which are

used together with a cartridge holder. For the installation of a preco-

lumn cartridge into the accompanying cartridge holder see the attached

leaflet.

The precolumn that is suitable for your separating column can be

found in the

Metrohm IC Column Catalog which can be obtained

from your local Metrohm agency, the data sheet accompanying your

separating column, the product information about separating columns

that can be found under

http://www.metrohm.com

, ion chromatogra-

phy products, or let your agency advise you directly.

New IC precolumns are normally filled with solution and sealed at both

ends. Before the precolumn is installed in the system, it must be en-

sured that this solution is freely miscible with the eluent used (check

manufacturer's specifications).