4 853 co2 suppressor – tubing connection, 10 putting into operation, 853 co – Metrohm 861 Advanced Compact IC User Manual
Page 53: Suppressor – tubing connection, Putting into operation
2.10 Putting into operation
861 Advanced Compact IC / Instructions for Use 8.861.1033
43
1 Electrical connection 853 – 861
•
Use the 6.2143.230 Cable to connect the Remote input of the
853 CO
2
Suppressor to Remote interface 21 of the 861 Ad-
vanced Compact IC.
•
On the rear of the Remote connector further devices can be
connected via Remote connections.
In the PC program «IC Net» (from Version 2.3 SR2) the 853 CO
2
Sup-
pressor can be controlled by the system driver of the 861 (Version
2.861.0040) (see «IC Net» Instructions for Use, Section 6.26).
2.9.4 853
CO
2
Suppressor – Tubing connection
The 853 CO
2
Suppressor is included between suppressor module
«MSM II» 46 and detector block 45.
1 Connection suppressor module «MSM II» – 853
•
Connect outlet capillary 83 (marked with "Detector") of sup-
pressor Module «MSM II» 46 to the inlet
Eluent in of the 853
CO
2
Suppressor using a 6.2744.010 Compression fitting.
2 Connection Detector block – 853
•
Connect inlet capillary 44 of detector block 45 to the outlet
Eluent out of the 853 CO
2
Suppressor using a 6.2744.010
Compression fitting.
2.10 Putting
into
operation
2.10.1 Putting into operation without suppressor module «MSM II»
Before sample solutions can be injected at the 2.861.0010 Compact
IC (without suppressor), the entire system must be tested for leaks and
then conditioned with eluent until the baseline is stable. Proceed as fol-
lows:
When inserting the column always make sure that it is inserted cor-
rectly in the flow direction shown on the sticker (the arrow must point
in the flow direction).
1 Open and connect system
•
Start the «IC Net» PC program, if it has not already been
started (see Section 2.5.3).
•
Select
File / Open / System
in the main window. Open the sys-
tem file created in Section 2.5.4 for the 861 Advanced Com-
pact IC.