Metrohm tiamo 2.2 Manual User Manual
Page 1293

■■■■■■■■■■■■■■■■■■■■■■
6 Configuration
tiamo 2.2
■■■■■■■■
1279
Maximum Dosing/Filling rate for exchange unit as a function of the
cylinder volume:
Cylinder volume
Maximum rate
1 mL
3.0 mL/min
5 mL
15.0 mL/min
10 mL
30.0 mL/min
20 mL
60.0 mL/min
50 mL
150.0 mL/min
Note
Enter lower rates for high-viscosity liquids.
Tubing parameters
Defines the length and diameter of the tubing connected to the exchange
unit.
Note
Default values have already been entered for the tubing parameters; these
correspond to the dimensions of the standard tubing supplied. If you do
not make any alterations to the tubing connections then you do not need
to modify the tubing parameters. For information on length and diameter
of other tubing see Products/Accessories and columns under
Dosing tip
Length
Length of the tubing connecting the cock to the dosing tip.
Range
0.0 ... 999.9 cm
Default value
40.0 cm
Diameter
Diameter of the tubing connecting the cock to the dosing tip.
Range
0.0 ... 9.9 mm
Default value
2.0 mm
Cylinder