3 meas ref - measuring parameters, Measuring parameters – Metrohm tiamo 2.2 Manual User Manual
Page 996

5.6 Commands
■■■■■■■■■■■■■■■■■■■■■■
982
■■■■■■■■
tiamo 2.2
5.6.4.12.3
MEAS Ref - Measuring parameters
Tab: Method
▶ MEAS Ref ▶ Properties... ▶ Measuring parameters
Command name
Name of the command.
Input
25 characters
The parameters for the measuring procedure can be defined on this tab.
Start wavelength
Lower limit of the spectrum.
Range
200.0 ... 1100.0 nm
Default value
400.0 nm
End wavelength
Upper limit of the spectrum.
Range
200.0 ... 1100.0 nm
Default value
1000.0 nm
Integration time
Integration time for spectrum recording.
Range
0.01 ... 600000 ms
Default value
6 ms
Averaged spectra
The number of spectra that are recorded and averaged.
Range
1 ... 10000
Default value
10
Smoothing
Number of neighboring pixels with which the value for each measurement
pixel is determined.
Range
0 ... 100 Pixels
Default value
0 Pixels