ALTANA BYK-mac i COLOR Manual User Manual
Page 32
31
Measure
After the standard has been measured, or when it is
already stored in the instrument, the name of the
checkzone to be measured first, appears on the left
side of the display.
The number of performed and predefined
measurements (0 of 2) appears in the upper right
corner.
Press the operate button to start the reading.
Once the number of measurements for the
checkzone is reached, a double audio signal is
heard.
The display shows the results of the measurement
and indicates that the measurement of the
checkzone is complete (e.g.2/2).
Pressing briefly the operate button allows the next
checkzone to appear in the display.
Once all checkzones have been measured, the
instrument returns to the Measure menu.
Pass/Fail Mode
In the software smart-process tolerances can be set
for each color. If Pass/Fail mode is selected in the
organizer and the measurements are within the
tolerance range, PASS is indicated in the display. If
not, FAIL.
To see which measurement value is out of tolerance,
you can scroll through the display pages with the
scroll wheel.
The appropriate value will be highlighted.