”Vac control”

• With pressure preselection, switches a pump or an in-line valve to

maintain that pressure in two-point control.

• With pressure preselection, controls a NT VARIO / VARIO-B pump to

maintain pinpoint control of that pressure.

• Coolant valve

”Auto mode” (only available with NT VARIO / VARIO-B pump)

• Provides fully automatic boiling point determination and adaptation

with pinpoint precision, and optimization of pumping speed with NT

VARIO / VARIO-B pumps.

• Coolant valve

’’Program’’

• Control pump or in-line valve based on time and pressure preselec-

tions, or ”Auto mode” (”Auto mode”

is only available with a NT VARIO

/ VARIO-B pump).

• Coolant valve

• Venting valve

”VACUULAN”

• Switches a pump or an in-line valve depending on preselected pres-

sure and time settings.

• Use continuous speed control to manage NT VARIO / VARIO-B pumps,

based on pressure and time preselections.

• Coolant valve

’’Configuration’’ (also accessible by pressing the selection knob while

the start display is shown)

Preselections for

• Adjustment of the pressure transducer

• Interface RS-232

• Sensors (configuration and switching between several sensors)

• Display (brightness and contrast of the display, language, sound)

• Autostart (automatic restart after power failure)

• Defaults (reset the controller to factory set values)