Speed (only for NT VARIO / VARIO-B pumps): The maximum motor

speed can be preselected (to control the pumping speed).

The ”Speed” is adjustable in a range of 1-100% or can be set to ”HI”.

The selection ”HI” (recommended) provides the optimum pumping

speed of the pump and therefore an optimum ultimate vacuum.

+

Maximum: An upper pressure limit can be preselected. The pump

switches off if the pressure limit is exceeded, e.g., at the end of suc-

tions or filtrations. The pressure limit is only active once the pressure

has gone below ”Maximum”.

The ”Maximum” is adjustable in a range of 794-1Torr (1059 - 1 mbar)

(at the least 1 Torr (mbar) higher than the ”Set vacuum”) and to ”Off”.

”Off” means that no ”Maximum” value is preset.

+

Delay: Delay” determines the time the coolant valve remains open af-

ter the process has been stopped. Determines also the time the pump

(only with VMS module and in-line valve) remains running after the

process has been stopped.

The ”Delay” is adjustable in a range of 1-300 minutes or can be set to

”Off” (”Off” means that the coolant valve closes immediately - and that

a pump with VMS module and in-line valve is switched off immediately

- when the process stops.).

+

Duration: Duration” determines the total process time since control

start.

A preset ”Duration” (process time) has no effect if the process is

stopped due to a preset ”Maximum” before ”Duration” is reached.

The ”Duration” is adjustable between 1-1440 minutes (24 h) or can be

set to ”Off”. ”Off” means that no endpoint of the process is defined.

Vac control

When selecting ”Graphic” the dis-

play shows a pressure vs. time

curve.

The timeline in the diagram

adapts automatically to the pro-

cess time.
+

Press the selection knob twice

to return to the standard dis-

play.