alternatively:

Fully automatic determination and adaptation of the boiling point

with NT VARIO / VARIO-B diaphragm pumps (recommended)

+

Select ”

Auto mode”.

+

Start process by pressing ”START/STOP” key.

+

The ”Auto mode” allows a fully automatic distillation even if the boiling

vacuum is unknown. In case of solvent mixtures the vacuum is reduced

until all solvents or volatile components are evaporated (with respect

to the ultimate vacuum of the pump and the bath temperature).

+

If ”Minimum” is set, the controller switches off the pump when the pre-

set vacuum value is reached. The ”Minimum” should be well below

the boiling vacuum of the solvent to be evaporated and above the ul-

timate vacuum attainable in the apparatus. If the ”Minimum” is set to

Auto”, the controller automatically switches off the pump at the end

of the evaporation (recommended for large solvent quantities only).

The setting of ”Duration” (process runtime) has no effect if the pump is

switched off due to ”Minimum” before ”Duration” is over.

+

When setting a value for ”Duration” the controller switches off the

pump when ”Duration” has passed even if a preset ”Minimum” is still

not reached.

+

If neither ”Minimum” nor ”Duration” is preset, pumping down has to be

finished by pressing the ”START/STOP” key.

+

Select function ”

VACUULAN”.

+

Set ”Set vacuum” to a pressure which can be reached reliably in the

vacuum network. Take account of the ultimate vacuum of the pump

and of the system’s leak rate in case of no vacuum demand.

+

Set ”Switch on” pressure appropriately to ensure sufficient vacuum for

all connected applications.

+

Set ”Delay” if necessary.

+

Start process by pressing ”START/STOP” key.