4.5 PWM 8 MODE: MEASURE AMPLITUDES


In this mode the vertices of the amplitudes of the incremental signals 1 and 2 are measured. The result
always refers to an individual signal period. With sinusoidal encoder signals (11µApp and 1Vpp) the positive
and the negative vertices are measured versus U

0

, with square-wave encoder signals (TTL and HTL) low

and high level are measured versus 0V.

In the table below the maximum measuring ranges are listed for the different interface boards:

Interface board:

11µApp

1Vpp

TTL

HTL

Maximum measuring range

33 µApp

1.66 Vpp

low: 0 – 2.5 V
high: 2.5 – 7.5 V

low: 0 – 7.5 V
high: 7.5 – 22.5 V


If the EXPERT MODE is active (see section 5) and the 11 µApp or 1Vpp interface board used, the encoder
supply voltage can be set in the mode MEASURE AMPLITUDES:

For this purpose the standard soft-key row has been expanded by the soft key

.







Soft key to switch to setting of encoder voltage



After pressing the soft key

the encoder voltage can be set:







Soft key to switch back to the standard
soft-key row

Display of encoder voltage

Soft keys to set the encoder voltage