- 54 -

TTL signal level

UH

2.5 V with

IH = 20 mA

UL

0.5 V with IL = 20 mA

Load capacity

IH

20 mA

IL

20 mA

CLoad

1000 pF

Switching times

Rise time:

t+

100 ns

Fall

time:

t

100 ns

The fault detection signal

UaS

indicates a malfunction, such as break of a supply line, lamp failure etc.

TTL square-wave signals can be transferred over cable lengths up to 300 m to the subsequent electronics.
For this purpose a supply voltage of 5 V

±

10% (for RAN 460: range between 10 and 30 V) must be ensured

at the linear or rotary encoder.
In the subsequent electronics TTL square-wave signals can be interpolated 100-fold max. by means of
phase-locked control loops.


Recommended input circuit of the subsequent electronics


Recommended differential line receivers
AM 26 LS 32
MC 3486
SN 75 ALS 193
SN 75 ALS 195

We recommend not to exceed the cable length related to the scanning frequency, since otherwise the
switching times resp. the edge gradient cannot be maintained.