TROUBLESHOOTING

DRO 200E

32

Incorrect Operation

If system positioning does not seem to be repeatable, the problem could be with the
machine tool or with the readout system.

Check linear scales.

Check each scale and reading head for proper installation.

Ensure that the mounting brackets are secure. Verify repeatability of the read-
ing head.

If the system seems to be displaying incorrect positions, check the following items.

Verify presets.

Make sure the numbers you enter are correct.

Verify tool.

Check that the correct electrode is being used. Check the electrode

for wear.

Verify the scale factor.

Ensure that the correct scale factor is being used.

Verify linear error compensation.

Make sure that the factors used for linear

error compensation are correct.

Diagnostics

Start Up Diagnostics

There are four diagnostic tests performed each time the readout is powered up.
These tests run without indication unless a problem is detected. If a problem is
detected, a message will be displayed at start up. The messages are as follows:

If you experience any of these error messages please contact your ACU-RITE dis-
tributor for assistance.

Operator-Performed Diagnostics

There are four operator-performed diagnostic tests to evaluate your readout. The
tests can be operated by holding the decimal key for two seconds. When the soft-
ware version is displayed, you are in the Diagnostics mode.

Test

Key Pad Test

EEPROM Test

RAM Test

ROM Test

Error Message

“KEY STUCK R(#) N(#) - PRESS CLEAR

MEMORY FAILURE [1] - PRESS CLEAR
MEMORY FAILURE [2] - PRESS CLEAR
MEMORY FAILURE [3] - PRESS CLEAR

CAUTION

If any memory failure is indicated the readout cannot
be relied on for correct operation. The readout should
be serviced immediately.